Hydraulic Power Unit Specifications
Landmark Test System Site Prep Guide System Component Specifications
19
Model Specific Specifications
Parameter 505.07 Specification 505.11 Specification
Pump/motor
Maximum Continuous
pressure
Maximum Flow
capacity
Motor rating
Line voltage starter configuration
21 MPa (3000 psi)
22.7 L/m (6 gpm) at 50 Hz
26.5 L/m (7 gpm) at 60 Hz
11.2 Kw (15 hp) at 50/60 Hz
Wye-Delta starter configuration
21 MPa (3000 psi)
41.6 L/m (11 gpm) at 50/60 Hz
18.6 Kw (25 hp) at 50/60 Hz
Water flow rating
(input temperature)
10.0°C (50°F)
15.5°C (60°F)
21.1°C (70°F)
26.7°C (80°F)
32.2°C (90°F)
Heat load (maximum)
3.8 L/m (1.0 gpm)
4.9 L/m (1.3 gpm)
6.1 L/m (1.6 gpm)
8.3 L/m (2.2 gpm)
15.9 L/m (4.2 gpm)
12.3 kW (42,000 Btu/hr)
7.2 L/m (1.9 gpm)
9.1 L/m (2.4 gpm)
12.1 L/m (3.2 gpm)
18.9 L/m (5.0 gpm)
49.2 L/m (13.0 gpm)
20.5 kW (70,000 Btu/hr)
Weight with maximum oil
174 L (46 gal))
450 kg (992 lb) 473 kg (1042 lb)