Model 494.16 VD/DUC Mezzanine Card
MTS FlexTest® Models 40/60/100/200 Controller Hardware Mezzanine Cards
163
4. Install the VD/DUC card(s) on the I/O carrier board.
5. (Optional) Install shunt-calibration-resistor plug assemblies on the front
panel of the I/O carrier board(s).
Note If you use MTS TEDS modules or MTS transducers with integrated
shunt-calibration resistors, you must insert a jumper plug (MTS part
number 100-188-097) into the socket for each transducer input
where you will use the integrated shunt-calibration resistor.
6. Install the I/O carrier board(s).
A. Plug the board into the VME card cage.
B. Connect the transducer cables to the I/O carrier board.
C. Connect the valve cable to the I/O carrier board.
7. After you have installed all boards, apply power to the system and use
the hardware-mapping and control software to configure the
conditioner and the valve driver.
8. Allow the system to warm up for an hour and then perform valve-
balance, calibration, and tuning procedures.