Mts FlexTest Models 200 Manuel d'utilisateur Page 266

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 344
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 265
MTS FlexTest® Models 40/60/100/200 Controller Hardware
Model 494.74 Two-Station HSM Interface Board
Digital I/O and Transition Boards
266
Cable specification To maintain EMC compliance, J3 Interlock cables must comply with the
following specifications:
Connector–15-contact, type D, male EMI connector.
Backshell–EMI metallized plastic or metal.
Cable–AWG and number of conductors as required. Braided shield with
shield connected to the metallized backshell at the chassis.
Jumper plug required If connector J3 is not used, you must install a jumper plug to maintain the
integrity of the interlocks. Use jumper plug part number 100-180-351 or
jumper pins: 1 and 4; 2 and 10; 9 and 12.
Vue de la page 265
1 2 ... 261 262 263 264 265 266 267 268 269 270 271 ... 343 344

Commentaires sur ces manuels

Pas de commentaire