MTS FlexTest® Models 40/60/100/200 Controller Hardware FlexTest Controller Configurations
49
Specifications–Model 494.04 Chassis
All equipment related to the controller should be connected to the same fused
power circuit.
Note Electrical connections must be made by qualified personnel and conform
to local codes and regulations. Local electrical codes supersede any
information found here.
Model 494.04 Chassis Specifications
P
ARAMETER SPECIFICATION
Input Voltage 100–240 V AC (single phase)
power factor corrected universal input
Input Frequency 50–60 Hz
Input Current 5 A at 100 V AC
3 A at 240 V AC
Facility Power Requirements Provide a dedicated circuit for the
chassis, computer, and monitor.
Input Surge <40 A
Insulation Over Voltage Category I
Pollution Degree 2
Weight Approximately 8.6 kg (19 lb)
Dimensions Width: 43 cm (17 in)
Height: 14 cm (5.5 in)
Depth: 44.5 cm (17.5 in)