Mts FlexTest Models 200 Manuel d'utilisateur Page 291

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 344
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 290
Optional Station Configurations
MTS FlexTest® Models 40/60/100/200 Controller Hardware Optional Station Configurations
291
Appendix B
Optional Station Configurations
About Optional Station Configurations
You can configure the Series 494 Chassis to support an optional six or eight
stations. When configuring your system for a six or eight stations, consider
the following:
Ensure that the .hwi file is correctly set for the desired multistation
configuration, especially the interlock and HSM board settings.
Power to each HSM is limited.
Cross-head interlocks with solenoid power are not supported.
AC input power must be at least 115 V AC.
Auxiliary power out of J49 on the Model 493.74 HSM Transition board
is not supported.
Six-Station Configurations
The six-station configuration can provide either six or eight channels of
control.
A typical 6-channel/6-station configuration requires 12 Digital
Universal Conditioners and 6 two-stage valve drivers.
A typical 8-channel/6-station configuration requires 16 Digital
Universal Conditioners and 8 two-stage valve drivers.
HSM power limits HSM power current is limited to 1.6 A per HSM.
Interlocks For 6-station configurations, the .hwi file must contain the line
INTERLOCKS=6. This line must be a discrete entry, not part of any other
hwi section.
Cross-head interlocks with solenoid power are not supported for 6-station
configurations.
Vue de la page 290
1 2 ... 286 287 288 289 290 291 292 293 294 295 296 ... 343 344

Commentaires sur ces manuels

Pas de commentaire